Paper/Archive/CVE-2024-9832

CVE-2024-9832

Baxter Life2000 Ventilation System — No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

Description · from the CVE record

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Affected

ProductAffectedFixed inSource
Life2000 Ventilation System06.08.00.00 and prior–CNA
life2000_ventilator_firmware≤ 06.08.00.00–CISA-ADP

Scores by source

Baxter (CNA)9.3CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CISA ADPSSVCExploitation: none · Automatable: no · Technical impact: total

References

refcisa.gov/news-events/ics-medical-advisories/icsma-24-319-01