CVE-2024-9832
Baxter Life2000 Ventilation System — No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
Description · from the CVE record
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Affected
ProductAffectedFixed inSource
Life2000 Ventilation System06.08.00.00 and prior–CNA
life2000_ventilator_firmware≤ 06.08.00.00–CISA-ADP
Scores by source
Baxter (CNA)9.3CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CISA ADPSSVCExploitation: none · Automatable: no · Technical impact: total