CVE-2024-48967
Baxter Life2000 Ventilation System — Life2000 ventilator and Service PC lack sufficient audit logging capabilities
Description · from the CVE record
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
Affected
ProductAffectedFixed inSource
Life2000 Ventilation System06.08.00.00 and prior–CNA
life2000_ventilator_firmware< 6.08.00.006.08.00.00CISA-ADP
Scores by source
Baxter (CNA)10.0CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CISA ADPSSVCExploitation: none · Automatable: yes · Technical impact: total