Paper/Archive/CVE-2024-48966

CVE-2024-48966

Baxter Life2000 Ventilation System — Life2000 service tools for test and calibration do not support user authentication

Description · from the CVE record

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Affected

ProductAffectedFixed inSource
Life2000 Ventilation System06.08.00.00 and prior–CNA
life2000_ventilator_firmware< 6.08.00.006.08.00.00CISA-ADP

Scores by source

Baxter (CNA)10.0CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CISA ADPSSVCExploitation: none · Automatable: yes · Technical impact: total

References

refcisa.gov/news-events/ics-medical-advisories/icsma-24-319-01