CVE-2023-31222
Medtronic Paceart Optima — Medtronic Paceart MSMQ Deserialization of Untrusted Data
Description · from the CVE record
Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing data to be deleted, stolen, or modified, or the Paceart Optima system being used for further network penetration via network connectivity.
Affected
ProductAffectedFixed inSource
Paceart OptimaVersions 1.11 and earlier–CNA
Scores by source
Medtronic (CNA)9.8CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CISA ADPSSVCExploitation: none · Automatable: yes · Technical impact: total