Paper/Archive/CVE-2020-6965

CVE-2020-6965

icscert — In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4…

Description · from the CVE record

In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, a vulnerability in the software update mechanism allows an authenticated attacker to upload arbitrary files on the system through a crafted update package.

Affected

ProductAffectedFixed inSource
GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 MonitorsApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3,CARESCAPE Central Station,v1.X,CARESCAPE Central Station,v2.X,B450,v2.X,B650,v1.X,B650,v2.X,B850,v1.X,B850,v2.X–CNA

Scores by source

NVD9.9CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H

References

refus-cert.gov/ics/advisories/icsma-20-023-01