Paper/Archive/CVE-2020-12041

CVE-2020-12041

icscert — The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored o…

Description · from the CVE record

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Affected

ProductAffectedFixed inSource
Baxter Sigma Spectrum Infusion PumpsSigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum LVP v8.x w/Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24–CNA

Scores by source

NVD9.4CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:H

References

refus-cert.gov/ics/advisories/icsma-20-170-04